XING Chengxia, LI Yongli, JIN Xuliang, et al. STUDY ON ACCURACY OF ON-LINE MONITOR FOR SODIUM CONTAINED IN STEAM OF THERMAL POWER UNITS[J]. Thermal power generation, 2012, 41(4): 42-45.
XING Chengxia, LI Yongli, JIN Xuliang, et al. STUDY ON ACCURACY OF ON-LINE MONITOR FOR SODIUM CONTAINED IN STEAM OF THERMAL POWER UNITS[J]. Thermal power generation, 2012, 41(4): 42-45.DOI:
the accuracy of sodium ions contained in steam measured by using second-order differential flame spectroscopic method
ion chromatographic method
and on-line sodium monitor method has been obtained
at the same time
the measured results by using on-line monitors manufactured from Orion
Swan
and Polymetron Corps being compared.Results show that the on-line sodium monitors can satisfy the requirement for accurately measuring the trace sodium ion contained in steam under condition of ensuring the optimal work condition for said monitors.The measuring accuracy of on-line sodium monitors is affected by the electrode behavior
the flow speed of liquid
the concentration of alkalination agent
and the external interference etc.
hence
cleaning
maintenance
and conservation for said monitors should be periodically carried out.