NIU Chaoqun, WANG Qiuliang, LI Yi, et al. Numerical Computation Methods for Evaluating the Electromagnetic Safety of Magnetic Resonance Imaging System[J]. 2017, 43(8): 2442-2453.
NIU Chaoqun, WANG Qiuliang, LI Yi, et al. Numerical Computation Methods for Evaluating the Electromagnetic Safety of Magnetic Resonance Imaging System[J]. 2017, 43(8): 2442-2453.DOI:
with the development of magnetic resonance imaging(MRI) technology
increasingly attention has been paid to electromagnetic safety of MRI. To summarize the researches in recent years
we introduce the biological effects and safety standards of three different kinds of electromagnetic fields(static magnetic field
gradient field
and radio frequency field) produced by MRI equipment firstly. On the basis of this
the numerical computing methods used for evaluating the electromagnetic safety of MRI are emphatically introduced
in which the human numerical model and numerical algorithm are studied in detail and evaluated for both accuracy and efficiency. The results of research show that when the electromagnetic fields generated by MRI exceed a certain limit
they will create significant biological effects
leading to the discomfort reaction of body. The current body voxel model has high resolution and can reflect the real human body to some extent. However
the general model limits the full advantages of high field MRI equipment
and developing a specific voxel model is an inevitable trend. The existing numerical algorithms used to evaluate the MRI electromagnetic safety include finite element method
moment method
and time domain finite difference method
as well as quasi-static approximation methods and mixed numerical calculation methods. These methods can accurately predict the electromagnetic field value in human tissue up to a point
but can not take into account accuracy and efficiency simultaneously
still need to further improved. The existing safety standards are not yet complete and should add relevant safety standards about open MRI
PET/MRI
LINAC/MRI and other new MRI system. The results can provide references for MRI designers
help to complement the existing MRI safety standards
then to simultaneously improve system performance and ensure electromagnetic safety.