Effect of fly ash on measurement accuracy of SO2 electrochemical sensor in thermal power plant
摘要
针对监测火电厂周围空气污染物时,SO_2电化学气体传感器出现测量不准和使用寿命较短等问题,以山西省朔州市某煤矸石发电厂飞灰为研究对象,在分析飞灰形貌特征的基础上,通过理论计算得到飞灰沉降速度和不同时间沉积量,通过实验测试飞灰对SO_2传感器测量误差的影响。研究结果表明:飞灰直径分布范围4~10μm,飞灰沉降速度2.25×10–3 m/s
Aiming at solving the problems of inaccurate measurement and short service life of SO_2 electrochemical gas sensor in monitoring air pollutants around thermal power plants
by taking the fly ash from a gangue-fired power plant in Shuozhou
Shanxi Province as the research object
and based on the analysis of the morphological characteristics of fly ash
the settling speed of fly ash and the amount of deposition at different times were obtained through theoretical calculation
and the influence of fly ash on the measurement error of SO_2 sensor was tested by experiments. The results show that
the diameter distribution range of the fly ash is 4~10 μm
the settling velocity of the fly ash is 2.25×10-3 m/s and the deposition amount in one day is 1.55 mg. The deposition time of fly ash will affect the measurement error of the sensor
the longer the deposition time
the greater the measurement deviation of the sensor. The reason is that the deposition of fly ash leads to the reduction of gas diffusion area s and diffusion coefficient K of the electrochemical sensor. Under the condition of fly ash deposition
the greater the gas mass concentration
the faster the gas reaches the working electrode area
the higher the effective SO_2 concentration detected in the limited response time
and the smaller the measurement error of the sensor. When the SO_2 electrochemical sensor is used in the environment with a large amount of fly ash
it is recommended to install a filter membrane with pore diameter less than 4 μm to prevent dust and clean the deposited fly ash regularly.