脑瘤是影响国民健康状况的重大难题。为了确定脑瘤的发展程度,以确定下一步治疗方案,通常需要对脑瘤组织进行穿刺活检手术。由于核磁共振成像(Magnetic Resonance Imaging
MRI)对软组织有更好的分辨率,常用来检测脑部肿瘤。因此,针对MRI兼容的机器人的相关研究是非常必要的。本文基于一种核磁共振兼容的液压驱动穿刺手术机器人,根据液压连通器的原理,推导了该机器人的运动学模型,并基于流体力学的相关理论,得到了该机器人的动力学模型。为了精确控制所设计的机器人系统,根据
$ {H}_{\infty } $
<
img text_id='M1' class='formula-img' style='display:none;' src='gdgydxxb-40-1-68_M1.jpg'>
控制理论设计了该液压驱动系统的状态反馈
$ {H}_{\infty } $
<
img text_id='M1' class='formula-img' style='display:none;' src='gdgydxxb-40-1-68_M1.jpg'>
控制率,使机器人可以快速、稳定地跟踪目标信号。最后,通过实验研究,该机器人系统的综合定位精度为0.56 mm,在
x
轴、
y
轴、
z
轴、俯仰轴和横滚轴上的平均定位精度分别为0.41 mm、0.6 mm、0.67 mm、0.886°和1.17°。研究结果验证了机器人辅助定位穿刺针的性能,所建立的动力学模型和控制方法对穿刺机器人的控制算法研究有一定的参考价值。
Brain tumors are a major problem affecting the health status of the nation. To determine the extent of brain tumor in order to determine the nex
t step in treatment
a puncture biopsy procedure of brain tumor tissue is often required. Magnetic resonance imaging (MRI) is commonly used to detect brain tumors due to its better soft-tissue resolution. Therefore
research related to MRI-compatible robots is necessary. Based on an MRI-compatible hydraulically driven puncture surgery robot
the kinematic model of the robot is derived based on the principle of hydraulic linker
and the dynamic model of the robot is obtained based on the relevant theory of fluid dynamics. In order to realize the accurate control of the designed robot system
the state feedback
${H}_{\infty } $
<
img text_id='Z-20221130165923' class='formula-img' style='display:none;' src='gdgydxxb-40-1-68_Z-20221130165923.jpg'>
control rate of this hydraulically driven system is designed according to the
${H}_{\infty } $
<
img text_id='Z-20221029143922' class='formula-img' style='display:none;' src='gdgydxxb-40-1-68_Z-20221029143922.jpg'>
control theory
which enables the robot to track the target signal quickly and stably. Finally
the average positioning accuracy of the designed robot system in
x
-axis
y
-axis
z
-axis
pitch-axis and roll-axis are obtained through experimental studies
and they are
respectively
0.41 mm
0.6 mm
0.67 mm
0.886° and 1.17°. Experimental results verify the performance of robot-assisted positioning puncture needles
and the dynamic model and control method of the robot have been given
which provide certain reference value for the research of the control algorithm of puncture robots.
KHIEWVAN B, TORIGIAN D A, EMAMZADEHFARD S, et al . An update on the role of PET/CT and PET/MRI in ovarian cancer [J ] . European Journal of Nuclear Medicine Molecular Imaging, 2017, 44(6): 1079-1091.
MOREIRA P, BOSKMA K J, MISRA S. Towards MRI-guided flexible needle steering using fiber Bragg grating-based tip tracking[C]// 2017 IEEE International Conference on Robotics and Automation (ICRA) . Singapore: IEEE, 2017: 4849-4854.
BALL M W, ROSS A E, GHABILI K, et al . Safety and feasibility of direct magnetic resonance imaging-guided transperineal prostate biopsy using a novel magnetic resonance imaging-safe robotic device [J ] . Urology, 2017, 109: 216-221.
ZHANG T X, WEN Y S, LIU Y H. Developing a parallel robot for MRI-guided breast intervention [J]. IEEE Transactions on Medical Robotics and Bionics, 2020, 2(1): 17-27.
GROENHUIS V, SIEPEL F J, VELTMAN J, et al. Design and characterization of Stormram 4: an MRI-compatible robotic system for breast biopsy[C]// 2017 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS). Vancouver: IEEE, 2017: 928-933.
HE Z, DONG Z, FANG G, et al . Design of a percutaneous MRI-guided needle robot with soft fluid-driven actuator [J ] . IEEE Robotics and Automation Letters, 2020, 5(2): 2100-2107.
LI G, NIRAVKUMAR A P, LIU W Q, et al. A fully actuated body-mounted robotic assistant for MRI-guided low back pain injection[C]// 2020 IEEE International Conference on Robotics and Automation (ICRA) . Paris: IEEE, 2020: 5495-5501.
NIRAVKUMAR A P, CHRISTOPHER J N, PAULO A C, et al . An integrated robotic system for MRI-guided neuroablation: preclinical evaluation [J ] . IEEE Transactions on Biomedical Engineering, 2020, 67(10): 2990-2999.
GUO Z Y, DONG Z Y, KIT-HANG L, et al . Compact design of a hydraulic driving robot for int ra-operative MRI-guided bilateral stereotactic neurosurgery [J ] . IEEE Robotics Automation Letters, 2018, 3(3): 2515-2522.
GYEONG HU K, NIRAVKUMAR P, YAN J W, et al. Shoulder-mounted robot for MRI-guided arthrography: clinically optimized system[C]// 2019 41st Annual International Conference of the IEEE Engineering in Medicine Biology Society (EMBC) . Berlin: IEEE, 2019: 1977-1980.
NIRAVKUMAR A P, YAN J W, DAVID L, et al. Body-mounted robot for image-guided percutaneous interventions: mechanical design and preliminary accuracy evaluation[C]// 2018 IEEE/RSJ International Conference on Intelligent Robots and Systems. Madrid: IEEE, 2018: 1443-1448.
NYCZ C J, GONDOKARYONO R, CARVALHO P, et al. Mechanical validation of an MRI compatible stereotactic neurosurgery robot in preparation for pre-clinical trials[C]// 2017 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) . Vancouver: IEEE, 2017: 875-881.
FRANCO E, BRUJIC D, REA M, et al . Needle-guiding robot for laser ablation of liver tumors under MRI guidance [J ] . IEEE/ASME transactions on mechatronics:A joint publication of the IEEE Industrial Electronics Society and the ASME Dynamic Systems and Control Division, 2016, 21(2): 931-944.
DONG Z Y, GUO Z Y, KIT-HANG L, et al . High-performance continuous hydraulic motor for MR safe robotic teleoperation [J ] . IEEE Robotics Automation Letters, 2019, 4(2): 1964-1971.
FARIMANI F S, MISRA S. Introducing pneuact: parametrically-designed MRI-compatible pneumatic stepper actuator[C]// 2018 IEEE International Conference on Robotics and Automation (ICRA) . Brisbane: IEEE, 2018: 4523-4536.
FRISHMAN S, KIGHT A, PIROZZI I, et al . Enabling in-bore MRI-guided biopsies with force feedback [J ] . IEEE Transactions on Haptics, 2020, 13(1): 159-166.