杜玉晓, 杨其宇. 医用新型高精密恒流电刺激器系统的设计[J]. 广东工业大学学报, 2020, 37(6): 17-25.
Du Yu-xiao, Yang Qi-yu. The Design of a New Type Constant Current Clinical Stimulator with High Precision[J]. 2020, 37(6): 17-25.
杜玉晓, 杨其宇. 医用新型高精密恒流电刺激器系统的设计[J]. 广东工业大学学报, 2020, 37(6): 17-25. DOI: 10.12052/gdutxb.200079.
Du Yu-xiao, Yang Qi-yu. The Design of a New Type Constant Current Clinical Stimulator with High Precision[J]. 2020, 37(6): 17-25. DOI: 10.12052/gdutxb.200079.
提出一种新型医用高精密恒流电刺激器系统的整体方案
可作为诱发电位的电刺激器
也可用作功能式神经肌肉电刺激器。针对高精密恒流电刺激器对刺激电压高伏值和刺激电流高精确度的设计要求
电源设计部分采用推挽升压后再倍压整流得到一个平稳的360 V高压
同时电路中的12 bit D/A和运放组成的恒流源电路可以保证一个高质量的脉宽低于10 μs、最大100 mA恒定电流的产生。针对高精密恒流电刺激器的高人体安全系数的要求
对电路中的控制信号、反馈信号和通信信号采用光电隔离
并对供电电源采用DC-DC电源隔离屏蔽; 同时设计了过流保护电路、脉宽限制电路和脉宽检错电路
结合底层硬件保护和上层脉宽检错电路的双重安全保护措施
大大提高了刺激器的安全系数。系统硬件电路的仿真和实际测试结果证明该设计方案的可行性和可靠性可满足临床医疗检测的要求。
The purpose of this paper is to illustrate a new scheme for a new medical high-precision constant current electric stimulator system. Methods: According to the design requirements of stimulation voltage of high-voltage and stimulation current of high-precision
the power supply adopts push-pull boost and then double-voltage rectification to obtain a stable 360 V high voltage. The 12 bit D/A and op amp constant current source circuit in the circuit could ensure that a high-quality pulse width is less than 10 μs and a maximum constant current is 100 mA. As for the high human safety factor
the control signals
the control signals
the feedback signals and the communication signals in the circuit are isolated by digital isolation
and the power supply is shielded by DC-DC power supply. At the same time
the over-current protection circuit
the pulse width limiting circuit and the pulse width error detection circuit are designed. Results: According to the high-amplitude voltage
high-precision current and high-safety factor of the human body
a system can be designed
which can be used as an electric stimulator for evoked potential or as a functional neuromuscular electrical stimulator. Conclusion: The feasibility and reliability of the system are proved by the hardware circuit simulation diagram and the actual test results. Concurrently
the intensity
frequency and pulse width of the stimulation current reach higher precision
the electric stimulation evoked potential waveform is clear
and the noise is small. Significance: The scheme of this new type of medical high-precision constant current stimulator system has been proved to meet the requirements of clinical medical examination.
DING J, WEXLER A S, BINDER-MACLEOD S A. Apredictive fatigue model part I: predicting the effect of stimulation frequency and patternon fatigue [J]. IEEE Transactions on Neural Systems and Rehabilitation Engineering, 2002, 10(1): 48-58.
VERAART C, GRILL W M, MORTIMER J T. Selective control of muscl activation with a multipolar nerve cuff electrode [J]. IEEE Transactions on Biomedical Engineering, 1993, 40(1): 640-645.
NATHAN R, TAVI M. The influence of stimulation pulse frequency on the generation of joint moments in the upper limb [J]. IEEE Transactions on Biomedical Engineering, 1990, 37(3): 317-321.
LADENBAUER J, MINASSIAN K, HOFSTOETTER U S, et al . Stimulation of the human lumbar spinal cord with implanted and surface electrodes: a computer simulation study [J ] . IEEE Transactions on Neural Systems and Rehabilitation Engineering, 2010, 8(6): 637-645.
HUANG H, HE J P, HERMAN R, et al . Modulation effects of epidural spinal cord stimulationon muscle activities during walking [J ] . IEEE Transactions on Neural Systems and Rehabilitation Engineering, 2006, 14(1): 14-22.
GRAUPE D, KORDYLEWSKI H. Artificial neural network control of FES in paraplegics for patient responsive ambulation [J]. IEEE Transactions on Biomedical Engineering, 1995, 42(7): 699-707.
GEORGIOU J, TOUMAZOU C. A 126 μW Cochlear chip for a totally implantable system [J]. IEEE Journal Solid-State Circuits, 2005, 40(2): 430-443.
BHATTI P, WISE K. A 32-site 4-channel high-density electrode array for a Cochlear prosthesis [J]. IEEE Journal of Solid-State Circuits, 2006, 41(12): 2965-2973.
THEOGARAJAN L S. A low-power fully implantable 15-channel retinal stimulator chip [J]. IEEE Journal of Solid-State Circuits, 2008, 43(10): 143-145.
NOORSAL E, SOOKSOOD K, XU H C, et al . A neural stimulator frontend with high-voltage compliance and programmable pulse shape for epiretinal implants [J ] . IEEE Journal of Solid-State Circuits, 2012, 47(1): 244-256.
TOKUDA T, HIYAMA K, SAWAMURA S, et al . CMOS-based multichip networked flexible retinal stimulator designed for image-based retinal prosthesis [J ] . IEEE Transactions on Electron Devices, 2009, 56(11): 2577-2585.
HART R L, BHADRA N, MONTAGUE F W, et al . Design and testing of an advanced implantable neuroprosthesis with myoelectric control [J ] . IEEE Journal of Solid-State Circuits, 2011, 19(1): 45-53.
THURGOOD B K, WARREN D J, LEDBETTER N M, et al . A wireless integrated circuit for 100-channel charge-balanced neural stimulation [J ] . IEEE Transaction on Biomedical Circuit and Systems, 2009, 3(6): 405-413.
LIU X, DEMOSTHENOUS A, DONALDSON N. An integrated stimulator with DC-isolation and fine current control for implanted nerve tripoles [J]. IEEE Journal of Solid-StateCircuits, 2011, 46(7): 1701-1714.
CHIANG B, FRIDMAN G Y, CHENKAI D, et al . A micropower low-noise neural recording front-end circuit for epileptic seizure detection [J ] . IEEE Journal of Solid-State Circuits, 2011, 46(6): 1392-1405.
CHIANG B, FRIDMAN G Y, CHENKAI D, et al . Design and performance of a multichannel vestibular prosthesis that restores semicircular [J ] . IEEE Transactions on Neural Systems and Rehabilitation Engineering, 2011, 19(5): 45-53.
LANGLOIS P J, IOANNIS P. High-power integrated stimulator output stages with floating discharge over a wide voltage [J]. IEEE Transaction on Biomedical Circuit and Systems, 2010, 4(1): 39-48.
CHENG K W, LU Y, TONG K Y, et al . Development of a circuit for functional electrical stimulation [J ] . IEEE Transactions on Neural Systems and Rehabilitation Engineering, 2004, 12(1): 43-46.
BROSER P J, CHRISTOPH B. Hydraulic driven fast and precise nonmagnetic tactile stimulator for neurophysiological and MEG measurements [J]. IEEE Transactions on Biomedical Engineering, 2012, 59(10): 2852-2858.
PENG J, DU Y X, YANG Q Y, et al . Design of a new type of EEG visual evoked potential flash stimulator [J ] . Journal of Guangdong University of Technology, 2010, 27(3): 60-63.
DU Y X, TANG Z Y, ZHOU W, et al . Design and implementation of evoked potential system [J ] . Mechanical Electrical Engineering Technology, 2013, 42(6): 84-90.
0
Views
0
下载量
0
CSCD
Publicity Resources
Related Articles
Related Author
Related Institution
京公网安备11010802024621