PE clapboard was completely burned at high temperature
the residue was leached with sulfuric acid(1+4) ICP-AES was applied to the determination of Ni and Co element in the leaching solution. The analytical spectral lines of 231.604
228.615 nm were chosen respectively for determination of Ni and Co. Based on the data
working curves were prepared.Detection limits(3 s) of 0.004
0.002 mg/L were obtained for Ni
Co respectively. Precision was tested by analyzing 2 samples
giving values of RSD’s(n =6) in ranges of 1.21%-5.86%. Recovery was tested by standard addition method
giving values of recovery in the range of 96.1%-98.1%.