The test solution of copper slag were prepared in this trial by microwave digestion method
and the standard addition method was applied in vanadium-ammonium molybdate spectrophotometric color-imetric method
the standard addition calibration curve linear range was 2~10 μg/mL; Linear correlation coefficient was 0.999 7; Recoveries of spiked test solution was 95.04%~99.85%. This method was simple and rapid
both the decomposition rate was enhanced and the loss of ingredient was reduced
but also the coexistence interference of the sample test solution was effectively inhibited