Research on the self-calibration method of inductive voltage divider based on improved incremental calibration method
|更新时间:2025-12-06
|
Research on the self-calibration method of inductive voltage divider based on improved incremental calibration method
Vol. 62, Issue 6, (2025)
作者机构:
1. 中国计量大学 机电工程学院
2. 中国计量科学研究院 电磁计量科学研究所
作者简介:
基金信息:
DOI:
CLC:
Published:2025
稿件说明:
移动端阅览
JIN Tao, PAN Xianlin, SHI Zhaomin, et al. Research on the self-calibration method of inductive voltage divider based on improved incremental calibration method[J]. 2025, 62(6).
DOI:
JIN Tao, PAN Xianlin, SHI Zhaomin, et al. Research on the self-calibration method of inductive voltage divider based on improved incremental calibration method[J]. 2025, 62(6).DOI:
Research on the self-calibration method of inductive voltage divider based on improved incremental calibration method
This paper presents an improved method for calibrating inductive voltage divider by double-equilibrium reference potential incremental calibration method
which combines a follower that can work at high voltage with a differential voltage measurement device to solve the influence of leakage current on the voltage ratio of the inductive voltage divider during the self-calibration process; by connecting a follower at the output of the reference voltage divider
the current of the differential measurement loop will not flow into the reference voltage divider to ensure the stability of the reference potential. The developed inductive voltage divider is self-calibrated at 400 V with a standard uncertainty of 1.5×10-9 for the ratio and 2.3×10-9 rad for the phase.