DETERMINATION OF SOLUTION TEMPERATURE COEFFICIENT FOR PH MEASUREMENT OF NAOH TREATED BOILER WATER
|更新时间:2025-09-28
|
DETERMINATION OF SOLUTION TEMPERATURE COEFFICIENT FOR PH MEASUREMENT OF NAOH TREATED BOILER WATER
Thermal Power GenerationVol. 41, Issue 12, Pages: 81-83(2012)
作者机构:
西安热工研究院有限公司
作者简介:
基金信息:
DOI:
CLC:TK31
Published:2012
稿件说明:
移动端阅览
LI Junwan, SONG Jingxia Xi'an Thermal Power Research Institute Co.Ltd.. DETERMINATION OF SOLUTION TEMPERATURE COEFFICIENT FOR PH MEASUREMENT OF NAOH TREATED BOILER WATER[J]. Thermal power generation, 2012, 41(12): 81-83.
DOI:
LI Junwan, SONG Jingxia Xi'an Thermal Power Research Institute Co.Ltd.. DETERMINATION OF SOLUTION TEMPERATURE COEFFICIENT FOR PH MEASUREMENT OF NAOH TREATED BOILER WATER[J]. Thermal power generation, 2012, 41(12): 81-83.DOI:
DETERMINATION OF SOLUTION TEMPERATURE COEFFICIENT FOR PH MEASUREMENT OF NAOH TREATED BOILER WATER
The accuracy of pH measurement of boiler water with caustic treatment is badly influenced by temperature.Appropriate STC(Solution Temperature Coefficient)was needed in order to obtain accurate pH measurement when the sample temperature is much beyond 25 ℃(reference temperature).The theoretical derivation process of STC was complicated and the result uncertainty was great.Therefore
an efficient and reliable experimental method to determine the STC was presented
and the setting methods of STC value for common domestic online pH analyzers were introduced.Results showed that the STC of NaOH treated boiler water was between 0.026 pH/℃ and 0.028 pH/℃
which was close to that of weak electrolyte alkaline solution by calculation but quite different from that of pure water
common water and strong electrolyte solution.Proper selection of pH analyzer with STC compensation and setting function is important
and to ensure the accuracy of pH measurement for boiler water