Determination of trace anions in water and steam from power plants by capillary ion chromatography
|更新时间:2025-09-28
|
Determination of trace anions in water and steam from power plants by capillary ion chromatography
Thermal Power GenerationVol. 44, Issue 9, Pages: 19-25(2015)
作者机构:
1. 国网山东省电力公司电力科学研究院
2. 济南大学资源与环境学院
作者简介:
基金信息:
DOI:
CLC:TM621
Published:2015
稿件说明:
移动端阅览
ZHANG Yan, HAN Bingjun, TANG Xinjian, et al. Determination of trace anions in water and steam from power plants by capillary ion chromatography[J]. Thermal power generation, 2015, 44(9): 19-25.
DOI:
ZHANG Yan, HAN Bingjun, TANG Xinjian, et al. Determination of trace anions in water and steam from power plants by capillary ion chromatography[J]. Thermal power generation, 2015, 44(9): 19-25.DOI:
Determination of trace anions in water and steam from power plants by capillary ion chromatography
C2O2-4)in water and steam from boilers in power plants.The separation was achieved on IonPac AS18(250mm×0.4mm)capillary anions analytical column with optimized KOH gradient elution and 10μL injection at 0.010mL/min
and then each component's concentration was detected by suppressed conductivity.The results show that
under the chosen condition
the linear correlation coefficients of the above 10 anions were greater than 99.9%
the detection of limit with signal-to-noise ratio of 3were between 0.04μg/L and 0.8μg/L.After running for 96 hours
the relative standard deviations(RSDs)were still between0.03% and 5.81%
and the average recoveries ranged from 81%to 116%
indicating this method has better repeatability and accuracy.This detection method can be applied in numbers determination of trace anions and low-molecular-weight organic acids in water and steam of boilers in power plants.