Pulsed Magnetic Field Generator with High-variation Rate Based on Archimedes Spiral Coil
|更新时间:2025-12-08
|
Pulsed Magnetic Field Generator with High-variation Rate Based on Archimedes Spiral Coil
Vol. 43, Issue 2, Pages: 578-586(2017)
作者机构:
1. 重庆大学输配电装备及系统安全与新技术国家重点实验室
2. 国网天津市电力公司检修公司
3. 国网安徽省电力公司检修公司
作者简介:
基金信息:
DOI:
CLC:
Published Online:10 February 2017,
Published:2017
稿件说明:
移动端阅览
MI Yan, RUI Shaoqin, CHU Yidao, et al. Pulsed Magnetic Field Generator with High-variation Rate Based on Archimedes Spiral Coil[J]. 2017, 43(2): 578-586.
DOI:
MI Yan, RUI Shaoqin, CHU Yidao, et al. Pulsed Magnetic Field Generator with High-variation Rate Based on Archimedes Spiral Coil[J]. 2017, 43(2): 578-586.DOI:
Pulsed Magnetic Field Generator with High-variation Rate Based on Archimedes Spiral Coil
Pulsed magnetic field treatment with cancer has good application prospects depending on its advantages of non-invasive. One of the mechanisms for pulsed magnetic field causing biological effects is from the induced electric field and magnetic field induced current. In order to obtain a larger induced electric field to carry out the cell experiments of pulsed magnetic field treatment on tumor cells
we develop a pulsed magnetic field generator with a high-variation rate by Archimedes spiral coil and the self-breakdown air switch. And we calculate the magnetic field and the induced electric field around the Archimedes spiral coil. By using COMSOL Multiphysics software
magnetic field distribution characteristic of coil is optimizing designed. Then the performance of the generator is measured. Finally
the Archimedean spiral coil has 7 turns
and the outer diameter is 21 mm. The pulsed magnetic field parameters can be produced in the cell culture medium as follows: the magnetic induction magnitude of 0.92 T
the rise time of 78 ns
the pulse width of 1.8 μs. And rate of change of magnetic field reaches 1.1×107 T/s
induced electric field maximum is 55.9 kV/m
electric field uniform degree can achieve 84%
and the maximum of the induced current density reaches 1.7×104 A/m2. The results show that the pulsed magnetic field generator can meet the requirements of the experiment cells associated with pulsed magnetic therapy treating with tumors