Multi-parametric Finite Element Simulation on Thermal Effect of Liver Tumor Exposed to High-frequency Nanosecond Pulse Bursts
|更新时间:2025-12-08
|
Multi-parametric Finite Element Simulation on Thermal Effect of Liver Tumor Exposed to High-frequency Nanosecond Pulse Bursts
Vol. 43, Issue 8, Pages: 2477-2485(2017)
作者机构:
1. 重庆大学输配电装备及系统安全与新技术国家重点实验室
2. 国网天津市电力公司检修公司
作者简介:
基金信息:
DOI:
CLC:
Published Online:31 August 2017,
Published:2017
稿件说明:
移动端阅览
MI Yan, RUI Shaoqin, LI Chengxiang, et al. Multi-parametric Finite Element Simulation on Thermal Effect of Liver Tumor Exposed to High-frequency Nanosecond Pulse Bursts[J]. 2017, 43(8): 2477-2485.
DOI:
MI Yan, RUI Shaoqin, LI Chengxiang, et al. Multi-parametric Finite Element Simulation on Thermal Effect of Liver Tumor Exposed to High-frequency Nanosecond Pulse Bursts[J]. 2017, 43(8): 2477-2485.DOI:
Multi-parametric Finite Element Simulation on Thermal Effect of Liver Tumor Exposed to High-frequency Nanosecond Pulse Bursts
In order to combine characteristics of traditional irreversible electroporation and nanosesond pulses
and to take full advantages of non thermal treatment of pulsed electric fields
the high-frequency nanosecond pulsed bursts was introduced for tumor ablation
and the thermal effects of it were studied. The temperature and thermal damage of the liver and tumor tissue
which have been punctured one pair of needle electrodes
were simulated by the multi-parameter finite element method. The pulse voltage used in this study ranges from 1 to 4 k V
the pulse width ranges from 50 to 500 ns
and the repetition frequency is between 100 k Hz and 1 MHz. The total pulse length is 100 μs
and the pulse burst repetition frequency is 1 Hz. The temperature under the thermocouple coupling was calculated using the Pennes biological heat transfer equation
and the thermal damage was calculate according to the Arrhenius formula. The research results indicated that the maximum temperature of the tumor reached at 100 μs was 49.26 °C under the single pulse train; the maximum temperature was 40.4 ℃ at 1 s
and the cumulative degree of thermal damage after 1 s was only 0.001 6
it will not cause thermal damage. By parameter fitting
the maximum and final temperatures under any parameter combinations can be obtained. These results can provide theoretical basis of pulse parameter selection for future experimental researches.