Numerical Computation Methods for Evaluating the Electromagnetic Safety of Magnetic Resonance Imaging System
|更新时间:2025-12-08
|
Numerical Computation Methods for Evaluating the Electromagnetic Safety of Magnetic Resonance Imaging System
Vol. 43, Issue 8, Pages: 2442-2453(2017)
作者机构:
1. 中国科学院电工研究所
2. 中国科学院大学
作者简介:
基金信息:
DOI:
CLC:
Published Online:31 August 2017,
Published:2017
稿件说明:
移动端阅览
NIU Chaoqun, WANG Qiuliang, LI Yi, et al. Numerical Computation Methods for Evaluating the Electromagnetic Safety of Magnetic Resonance Imaging System[J]. 2017, 43(8): 2442-2453.
DOI:
NIU Chaoqun, WANG Qiuliang, LI Yi, et al. Numerical Computation Methods for Evaluating the Electromagnetic Safety of Magnetic Resonance Imaging System[J]. 2017, 43(8): 2442-2453.DOI:
Numerical Computation Methods for Evaluating the Electromagnetic Safety of Magnetic Resonance Imaging System
with the development of magnetic resonance imaging(MRI) technology
increasingly attention has been paid to electromagnetic safety of MRI. To summarize the researches in recent years
we introduce the biological effects and safety standards of three different kinds of electromagnetic fields(static magnetic field
gradient field
and radio frequency field) produced by MRI equipment firstly. On the basis of this
the numerical computing methods used for evaluating the electromagnetic safety of MRI are emphatically introduced
in which the human numerical model and numerical algorithm are studied in detail and evaluated for both accuracy and efficiency. The results of research show that when the electromagnetic fields generated by MRI exceed a certain limit
they will create significant biological effects
leading to the discomfort reaction of body. The current body voxel model has high resolution and can reflect the real human body to some extent. However
the general model limits the full advantages of high field MRI equipment
and developing a specific voxel model is an inevitable trend. The existing numerical algorithms used to evaluate the MRI electromagnetic safety include finite element method
moment method
and time domain finite difference method
as well as quasi-static approximation methods and mixed numerical calculation methods. These methods can accurately predict the electromagnetic field value in human tissue up to a point
but can not take into account accuracy and efficiency simultaneously
still need to further improved. The existing safety standards are not yet complete and should add relevant safety standards about open MRI
PET/MRI
LINAC/MRI and other new MRI system. The results can provide references for MRI designers
help to complement the existing MRI safety standards
then to simultaneously improve system performance and ensure electromagnetic safety.
Measurement of Vibration Characteristics of Power Cable Line Under Typical Laying Conditions
Health Effects of Extremely Low Frequency Electromagnetic Fields
Computation of Transient Electromagnetic Field Emission from Enclosure During GIS Disconnector Switching Operations
Research Progress in Transcranial Magnetic Stimulation Technology
Related Author
XIA Rong
RAO Wenbin
SHU Naiqiu
YANG Dan
ZHAO Jiankang
MENG Shaoxin
ZHANG Zhenpeng
BAO Jiali
Related Institution
China Electric Power Research Institute
School of Electrical Engineering, Wuhan University
Research Group of Biophysics and Medical Engineering, Zhejiang Provincial Key Laboratory of Bioelectromagnetics,School of Medicine, Zhejiang University
College of Electrical Engineering, Zhejiang University
State Key Laboratory of Power Grid Environmental Protection, Wuhan Branch of China Electric Power Research Institute