Detection of Trace Ethyne by Photoacoustic Spectroscopy based on Fiber-optic Cantilever Acoustic Sensor
|更新时间:2025-12-08
|
Detection of Trace Ethyne by Photoacoustic Spectroscopy based on Fiber-optic Cantilever Acoustic Sensor
Vol. 46, Issue 6, Pages: 1922-1928(2020)
作者机构:
1. 重庆大学输变电装备及系统安全与新技术国家重点实验室
2. 江苏国电南自海吉科技有限公司
作者简介:
基金信息:
DOI:
CLC:
Published Online:30 June 2020,
Published:2020
稿件说明:
移动端阅览
YANG Tianhe, CHEN Weigen, LI Zhijun. Detection of Trace Ethyne by Photoacoustic Spectroscopy based on Fiber-optic Cantilever Acoustic Sensor[J]. 2020, 46(6): 1922-1928.
DOI:
YANG Tianhe, CHEN Weigen, LI Zhijun. Detection of Trace Ethyne by Photoacoustic Spectroscopy based on Fiber-optic Cantilever Acoustic Sensor[J]. 2020, 46(6): 1922-1928.DOI:
Detection of Trace Ethyne by Photoacoustic Spectroscopy based on Fiber-optic Cantilever Acoustic Sensor
The concentration of trace ethyne gas dissolved in transformer oil can adequately reflect whether overheating and discharge failures occur inside the transformer. Photoacoustic spectroscopy is becoming an effective monitoring method for online gas detection. However
in traditional photoacoustic spectroscopy
electric microphones are usually used to detect the photoacoustic signals. Weak electrical signals are easily affected by the strong electromagnetic field environment around the transformer
which seriously affects the stability and sensitivity of trace gas detection. We researched the detection of trace ethyne by photoacoustic spectroscopy based on fiber-optic acoustic sensors.Combined with fiber laser power amplification technology
the excitation light power at an absorpti
on wavelength of ethyne was increased from 20 m W to 200 mW. The fiber-optic cantilever acoustic sensor head was made to match the resonance frequency of the resonance photoacoustic cell
which realized the dual resonance enhancement of the cantilever and the resonance photoacoustic cell. Wavelength modulation technology was used to detect the gas concentration with the second harmonic amplitude of the photoacoustic signal. The photoacoustic spectroscopy system based on fiber-optic acoustic sensors was built for trace ethyne detection
and stable detection of trace ethyne with good linearity